Views:

 

                   This Article is about: Technical questions regarding periodic maintenance and diagnostic tools for both.

                    the ii900 and ii910. 

  • On start-up does the unit run a self-diagnosis of its functionality to identify any system failures?  Yes.
  • Do the Microphone/Sensors/Unit ever require Calibration / Service? No calibration required. We offer a performance test through Fluke service which some people use annually to ensure product is working well. This includes a cleaning of the sensor head area as sometimes dust and debris builds up.
  • Will the unit display an error message if Microphones are blocked or malfunctioning without going into the settings? No, you need to use the self-test available on the ii900 in the settings area.
  • If there is a failure of the process of Testing the Microphones after repeat cleaning, please provide a Calibration Service Agent that can rectify units in each of the following ports LAX, SYD and MEL.  You need to send the instrument into the Fluke service but first call 1-888-993-5853 to setup an RMA.