Before starting your recording session with the 1736 or 1738 energy logger it is critical to setup the proper power configuration.  The most common of these configurations would be either the WYE configuration or the Delta configuration.

mceclip1.png

mceclip2.png

To setup the configuration push the meter key on the 1736 or 1738 and the display should look like the following image if you are in Energy Study mode. If you are in the Load Study mode will see only current is visible in your display.

mceclip3.png


Tap change configuration

Select topology from the list by using the arrow keys and select the enter key or tap the screen.


From the presented list of power configurations select the appropriate power configuration using the arrow keys.  Ideally if you have a neutral present it should be connected.  The Delta configuration does not use the neutral connection.  The WYE configuration will use the neutral.