Views:

1. Error FFFFE00 

Error FFFFE00 occurs when the measure board has come loose, and in order to fix it, the measure board needs to be locked.

mceclip0.png 

Remove the top panel and locate the measure board. Verify if the locking mechanism is in the unlocked position (see below). Then push the measure board down until it makes a secure contact with the

connector at the bottom.

mceclip1.png

Proceed to lock the measure board, and this should fix Error FFFFE00. Now put the top panel back, and turn the unit on to check if the error is gone.

mceclip2.png

2. Error CCFFE00

This error occurs when the measure board has gone bad, and it needs to be replaced. The part number for this board is 3699112.  

mceclip3.png

3. Error CFFFE00

This error occurs when the analog board has gone bad, and it needs to be replaced. The part number for the analog board is 3699101. 

mceclip4.png

4. Error F7FE00

This error occurs when a relay(s) is not working properly in the analog board. To fix this problem the analog board needs to be replaced, and the part number is 3699101.

mceclip5.png

5. Error 3FFFE00

a- Check cable connections

b- Replace analog board (part number: 3699101)

If the above doesn’t fix the problem:

c- Replace cables

d- Replace main power supply board (part number 3699101)

e- Replace digital board (part number 3699093)

mceclip6.png