Views:

The 6270A pressure controller (or 8270A/8370A) can be used to float a PG7601, PG7102, PG7202, PG9602 or PG9607 piston just like previous generation controllers such as a PPC4. Below are some of the necessary things to setup or things to check if having any issues that might not be obvious using the 6270A controller. The notes below refer to a 6270A controller but the same applies to 8270A and 8370A controllers. However 8270A and 8370A are available with 5 PMM (Pressure Measurement Module) slots or 2 PMM Slots. 6270A always has 5 PMM slots. 

 Hardware Setup:

  • The PG needs to have firmware version v3.1.4.308 or higher to work with the 6270A and thus also means cannot be used with older PG bases that do not have the new dedicated AMH connection. This might change in future firmware. It's recommended to have PG firmware v3.1.5.255 or higher. See this web page to update the PG firmware, PG7000/PG9000 Ver 3.x Embedded Software
  • PM600 modules have a slower response rate so PM200 and PM500 modules should be used when possible. 
  • Many PM200 modules are limited on how far they can measure below atmosphere. In situations where a PG7601 is going to be used sub-atmospheric, PM500 modules are normally preferable.
  • A PMM of the appropriate mode (absolute, gauge) needs to be installed in the 6270A. If using one PMM, install it in slot 5 (right-most slot to the left of the Pressure Control Module) of the 6270A. If using two PM Modules, install the higher pressure module in slot 5 and the lower pressure module in slot 4. Most PM600 modules have a lower limit that does not go down to -1 atm and/or low absolute pressure. PM500-BGxxx modules are most often preferred if low absolute or negative gauge pressures are needed because their lower limit is all the way down to -1 atm. PM200 and PM500 modules are limited to the inherent mode they are designed for unless a barometer module is added to the controller. For example, a PM500-BG700K module would only allow operation in gauge mode from -1 atm to 100 psi (700 kPa) if no barometer module is installed. With a barometer module like PM200-A100K installed, the absolute range would be 0 to 115 psia (800 kPa). 
  • The PMM that will be used with a piston and mass set should have a max pressure that is no more than the "Max Ratio" shown in the below table. 
  • Controller

    PMM Class

    Max Ratio

    6270A

    PM200

    4:1

    PM500/PM600

    8:1 (6:1 with 10 kPa/kg PC module)

    8270A

    PM200

    5:1

    PM500/PM600

    10:1 (6:1 with 10 kPa/kg PC module)

    8370A

    PM200/PM230

    5:1

    PM500/PM600/PM630

    10:1

  • Use a null-modem/cross cable to connect the 6270A COM1 to the PG COM3. This is a different cable than would be used with a PPC4.
  • Under menu Setup --> Instrument Setup --> Remote Port --> RS232 Setup,  the Remote IF needs to be set to Computer.
  • When AutoGen is turned on from the PG Terminal, it will change the 6270A to Fixed mode and use the highest range PMM installed. This should suffice for most pressures being set, similar to how an A7M utility sensor might be used on PPC4. If there are lower range modules that are preferred for use, they can be installed and you can switch to Auto mode or a different Fixed mode PM module if desired. However, note that it will change back to Fixed mode and the highest range module anytime AutoGen is initialized.
  • The 6270A will be put in PPC4 Emulation mode when AutoGen is turned on and this should not be changed. This the reason for the higher pressure module being in slot 5 and the lower pressure module being in slot 4.
  • The float target setting is not used... so the value that it is set to in the PG or in COMPASS does not matter. 

 COMPASS for Pressure Calibration Software Setup:

  • Need to use COMPASS v5 or higher
  • No need to setup 6270A as a Support Device in COMPASS because the Piston Gauge will control it. Ensure that the PG AutoFloat option is selected for Control in the COMPASS Test setup.

Recommended Configurations for Common Piston Gauge Setups:

PG7601, all Piston-Cylinders with either 35 kg or 38 kg Mass Sets

Model

Item Num

Description

6270A-NPT

4427608

MODULAR PRESSURE CONTROLLER CHASSIS, NPT MANIFOLD

PCM-STD-20M

4428630

PRESSURE CONTROL MODULE, STANDARD TURNDOWN

PM200-A100K

4363867

PRESSURE MEASUREMENT MODULE,100 KPA (15 PSI) ABS

PM500-BG1M

4869710

PRESSURE MEASUREMENT MODULE -0.1 to 1 MPa (-15 to 150 psi) GAUGE

PM500-BG10M

4869787

PRESSURE MEASUREMENT MODULE -0.1 to 10 MPa (-15 to 1500 psi) GAUGE

 

PG7202, 80 kg mass set, with PC-7200-100 and PC-7200-500 piston-cylinders

Model

Item Num

Description

8270A-2-NPT-PCM

5019166

MODULAR PRESSURE CONTROLLER, 2-BAY, NPT, W/PCM

PM200-G40M

4916470

PRESSURE MEASUREMENT MODULE, 40 MPA (6000 PSI) GAUGE

PM200-G10M

4363961

PRESSURE MEASUREMENT MODULE,0 TO 10 MPA (0 TO 1500 PSI) GAUGE

 

PG7202, 100 kg mass set, with PC-7200-200 and PC-7200-1 piston-cylinders

Model

Item Num

Description

8370A-2-PCM

5019215

HIGH-PRESSURE MODULAR CONTROLLER, 2-BAY, W/PCM

PM230-G100M

4916428

HIGH-PRESSURE MEASUREMENT MODULE, 100 MPA (15000 PSI) GAUGE

PM200-G20M

4363989

PRESSURE MEASUREMENT MODULE,0 TO 20 MPA (0 TO 3000 PSI) GAUGE

 

PG7202, 55 kg mass set, with all piston-cylinders

Model

Item Num

Description

8370A-2-PCM

5019215

HIGH-PRESSURE MODULAR CONTROLLER, 2-BAY, W/PCM

PM200-G14M

4363977

PRESSURE MEASUREMENT MODULE,0 TO 14 MPA (0 TO 2000 PSI) GAUGE

PM230-G100M

4916428

HIGH-PRESSURE MEASUREMENT MODULE, 100 MPA (15000 PSI) GAUGE