Note that the same barometer is used in PPC3, RPM4, PPCH, PPCH-G and newer models of PG7000/PG9000 with version 3 firmware and above from August 2008, SN 733 and above.
If you suspect the on-board barometer in a PPC4 is faulty, do the following:
Error code: Baro Sensor err: $01
PPC4 internal barometer PN
3126355 | 401898-06,ASSY,BAROMETER,100K NOVA (DWG U03634) |
View: SETTINGS>MORE>CALIBRATION>BAROMETER
At the bottom of the screen, please note the internal barometer reading, is it changing if you put your finger on the ATM port in back, or is it stuck at a number? If stuck on a number or not reading, the problem is likely the barometer.
1. Using the manual, remove the cover.
2. Using the figure below, remove and replace barometer assembly. This is a plug-n-play function.
ure below, remove the existing barometer module