Install software (free) download at https://us.flukecal.com/literature/software-downloads/software/io-toolkit-v10


Connect 1523-24 unit to the computer and power on with probe attached. Ensure serial port is on as software will attempt to auto detect the attached instrument.


Launch Software.

mceclip1.png
Click on the probe settings button.

mceclip2.png
Choose the correct connector type.

mceclip0.png
Choose/enter the date for Probe Type, Serial#, Conversion Type, Cal date.

Then enter the parameters for the TPW, A, B, C, D, A4, B4, coefficients.

Scroll the bar down and then enter the min/max temperatures for the probe.

Click the upload button. Software will ask for pass code (1234). Parameters will then upload.