This intermittent error is a known issue in MET/CAL version 8. The Fluke Calibration software team is troubleshooting the source of the problem, and we are looking for more information from MET/CAL users. If you encounter error E2372, please respond to this article with a comment, describing the specific steps you took before reaching this error, providing as much detail as possible.
E2372: Internal error in module "win": Font (1) has not been initialized.
To resolve this error, please follow these steps:
- Open MET/CAL Run Time.
- Click on Configure > Default Font.
- Change to a different font, click OK.
- Click on Configure > Default Font.
- Change to your original font (or the default: Courier New), click OK.
Default Font is typically the only one affected. Repeat steps 2-5 for the Display Font and Picture Font, if necessary.