Views:

Tech Notes

          

--------------------------------------------------------------------------------------------------------------------------------

Title: 9133 will not heat after I switched the voltage from 110V to 230V

Created: 9 Apr 2012

Last Revised: 9 Apr 2012

--------------------------------------------------------------------------------------------------------------------------------

 

 

Applies To: 9133 IR Calibrator

 

Problem Description: The 9133 IR Calibrator does not heat after the voltage was switched from 110V to 230V.

 

Resolution/Work Around:  The voltage on the 9133 IR Calibrator must have the ~ showing through the window.  If the sticker with the ~ is missing, then the voltage must be in the 110V position.  If 230V is showing, the unit will not heat.  Place the 110V in the window and connect 230V to the 9133. 

 

The 9133 should have come with a sticker over the 110V and 230V writing.  The 9133 has switchable power supplies.   Just make sure the correct fuses are in place.