Views:

After migrating your data from METTRACK into METTEAM, all of your procedures will be referenced in work orders and associated with your types but the PXE files themselves will not have been uploaded into METTEAM and associated with the Data Sheet field on the Procedure record. To do this, we have to go through the process of uploading them into METTEAM via METCAL Editor.

  1. Login to METCAL Editor
  2. Click the Tools menu and select "Upload Existing PXEs to MET/TEAM"
    mceclip2.png
  3. Navigate to where you PXE files are stored
  4. You will get a pop-up window showing the progress of the upload
    mceclip0.png
  5. You may get notifications if two PXEs appear to be associated with the same procedure asking you to select the correct one
  6. Once the upload is complete, you will receive a dialog box telling you that the process is finished along with the location of the log file
    mceclip1.png

Errors

Error: Could not load file or assembly 'GuidHelper, Version=3.0.0.3..."

Please note: There may be a bug in this process for certain versions of MET/TEAM.  If you get the following error:

 

The Editor is missing a library that is needed for this process.

If you experience this error, the solution is to copy the GuidHelper.dll file from the Runtime folder (C:\Program Files (x86)\Fluke\METCAL) to the Editor folder (C:\Program Files (x86)\Fluke\METCAL Editor\Extensions\METCAL.Editor.Project).  

Once the file is copied, restart the Editor.  The PXE file upload should work correctly.

Error: Hwnd of zero is not valid.

You may receive the following error while importing PXEs:

mceclip0.png

Workaround: Temporary disable Accuracy Check in METCAL Runtime on the workstation

  1. Login to METCAL Runtime
  2. Go to the Configure tab
  3. On the left, select Metrology
  4. Uncheck Accuracy Check
    mceclip1.png
  5. Click Apply
  6. Log back into METCAL Editor and go through the process of uploading your PXEs
  7. Log back into METCAL Runtime and check Accuracy Check again

For more information contact softwaresupport@flukecal.com