| Model Number | Specified AC Amperage Range | Meter Knob Position | Specified DC Amperage Range | Meter Knob Setting |
| Fluke-15B | 1A to 600A AC | Volts AC use mV range up to 400A, and Volts AC 4V range for 400A to 600A | 1A to 1000A DC | Volts DC mV range up to 400A, and Volts DC 4V range for 400A to 1000A |
| Fluke-17B | 1A to 600A AC | Volts AC use mV range up to 400A, and Volts AC use 4V range for 400-600A | 1A to 1000A DC | Volts DC mV range up to 400A, and Volts DC 4V range for 400A to 1000A |
| Fluke-15B Max | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 400A-1000A |
| Fluke-17B Max | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 400A-1000A |
| Fluke-15B+ | 4A to 600A AC | AC/DC mV toggle yellow button up to 400A, and Volts AC 4V range for 400A to 600A | 1A to 1000A DC | AC/DC mV toggle yellow button up to 400A, and Volts DC use 4V range for 400A-1000A |
| Fluke-17B+ | 4A to 600A AC | AC/DC mV toggle yellow button up to 400A, and Volts AC use 4V range for 400A to 600A | 1A to 1000A DC | AC/DC mV toggle yellow button up to 400A, and Volts DC use 4V range for 400A-1000A |
| Fluke-18B+ | 4A to 600A AC | AC/DC mV toggle yellow button up to 400A, and Volts AC use 4V range for 400A to 600A | 1A to 1000A DC | AC/DC mV toggle yellow button up to 400A, and Volts DC use 4V range for 400A-1000A |
| | | | | |
| Fluke-27-II | 1A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-28-II | 18A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-28-II EX | 18A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-77-4 | 10A to 600A AC | Volts AC use 6V range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-83-5 | 1A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-87-5 | 18A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-87-5 MAX | 18A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-88-5 | 1A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-101 | 6A to 600A AC | Volts AC mV range | 10A to 1000A DC | Volts DC use 6V range |
| Fluke-106 | 6A to 600A AC | Volts AC mV range | 10A to 1000A DC | Volts DC use 6V range |
| Fluke-107 | 6A to 600A AC | Volts AC mV range | 10A to 1000A DC | Volts DC use 6V range |
| | | | | |
| Fluke-113 | 60A to 600A AC | Volts AC 6V range | 60A to 1000A DC | Volts DC 6V range |
| Fluke-114 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-115 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-116 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-117 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-175 | 30A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-177 | 30A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-179 | 30A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-233 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-279 | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-287 | 1A to 600A AC | Volts AC mV range up to 500A, and Volts AC use 5V range for 500A to 600A | 1A to 1000A DC | Volts DC mV range up to 500A, and Volts DC use 5V range for 500A to 1000A |
| Fluke-289 | 1A to 600A AC | Volts AC mV range up to 500A, and Volts AC use 5V range for 500A to 600A | 1A to 1000A DC | Volts DC mV range up to 500A, and Volts DC use 5V range for 500A to 1000A |
| | | | | |
| Fluke-787 | 20A to 600A AC | Volts AC use mV range up to 400A, and Volts AC use 4V range from 400A to 600A | 1A to 1000A DC | Volts DC mV range up to 400A, and Volts DC use 4V range for 400A to 1000A |
| Fluke-787B | 20A to 600A AC | Volts AC use mV range up to 400A, and Volts AC use 4V range from 400A to 600A | 1A to 1000A DC | Volts DC mV range up to 400A, and Volts DC use 4V range for 400A to 1000A |
| Fluke-789 | 20A to 600A AC | Volts AC use mV range up to 400A, and Volts AC use 4V range from 400A to 600A | 1A to 1000A DC | Volts DC mV range up to 400A, and Volts DC use 4V range for 400A to 1000A |
| | | | | |
| Fluke-1577 | 30A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| Fluke-1587 FC | 30A to 600A AC | Volts AC use mV range | 1A to 1000A DC | Volts DC mV range up to 600A, and Volts DC use 6V range for 600A to 1000A |
| | | | | |
| Fluke-3000 FC | 6A to 600A AC | AC/DC mV toggle yellow button | 1A to 1000A DC | AC/DC mV toggle yellow button up to 600A, and Volts DC use 6V range for 600A to 1000A |