Views:

Q: Is a Pyrex window suitable for Infrared energy transmission?

A: The Pyrex window is just boron-doped high temp glass, it has no IR transmission above 3um and also is bad with ratio sensors, but certainly a no-go for LT (8-14 um) sensors. It is also not-suited to ratio or two-color sensors due to a "kink" in the transmission curves at the E1R or EF1R wavelenths.

 

Additional info:

 

BK7 is a form of borosilicate crown-glass as is Pyrex (more doping with Boron to achieve high temperature thermal shock resistance).

 

MT sensors choose Sapphire or Calcium Fluoride -sapphire is more rugged, Calcium Fluoride has a lower cost.

 

Data from Crystran in the UK for typical IR window materials:

 

G5, MT, P7. 3M, 2M, 1M, HT

1R, 2R, 1M, 2M, 3M, MT, CO2, CO, NOx, HT, P3

 

1R, 2R, 1M, 2M, 3M, HT

 

1R, 2R, 1M, 2M, 3M, HT


All sensors

 

1R, 2R, 1M, 2M, 3M, HT, P3


 

All sensors


P7 LT (with care)