This Article is about details related to the 17XX-CLAMP-ADPT-1V and 17XX-CLAMP-ADP-2V8.

Item 5363472 17XX-CLAMP-ADPT-1V, 17XX BNC ADAPTER FOR CLAMPS 1mV/A, MAX. 1V.

Item 5363485 17XX-CLAMP-ADPT-2V8, 17XX BNC ADAPTER FOR CLAMPS 1mV/A, MAX. 2.8V.


The first adapter is designed for use for current clamps with an output maximum value of 1V AC/DC.
The second is designed for use with current clamps with an output maximum of 2.8V AC/DC.
The 17xx Clamp Adapter is an accessory for the 17xx-series of Power Quality Analyzers.

Use the adapter to connect the current clamp with the BNC connector to the 17xx Analyzer.

This adapter is ready-to-use for all clamps with mV/A output. Select the adapter version depending on the voltage output of the clamp.

DC current is only supported by the 177x power quality analyzers.

Maximum 1V: Supports clamps with a range from 0.2 ARMS to 1000 ARMS or up to ±1500 A DC.
Maximum 2.8V: Supports clamps with a range from 4ARMS to 2800ARMS or up to ±4000 A DC.
 

A range of Fluke devices have been tested with the instrument; this information is included in the product instructions.

The information includes the correct current ratio to be set in the instrument along with the maximum output voltage.

Users should consult this table before ordering the adapter (and current clamp).

Below are details of the supported clamps including ratio settings and recommended product for each Fluke current clamp: