Applies To:  All large refrigerated baths

Problem Description:   

In an effort to make Fluke products RoHS compliant we have changed from the old manually adjusted version of this line monitor, that was not RoHS compliant, to a new digitally adjusted version, that is RoHS compliant.

Resolution/Work Around:   

The settings should be as follows:

Setup

Voltage: 115/230 (depending on model)

Time Delay: 120 seconds

Over/under voltage: 12%

Control mode: off

Response time: 5 seconds

Fault 1-5: [ignore these, they can be anything, it’s just a record of previous cutouts]