Tech Notes
--------------------------------------------------------------------------------------------------------------------------------
Title: Factory PID settings for my Metrology Well
Created: 9 Jan 2013
Last Revised:
--------------------------------------------------------------------------------------------------------------------------------
Applies To: Various Metrology Wells
Problem Description: Unit will not maintain or reach temperature. PID is wrong.
Resolution/Work Around: Set the PID back to factory default.
4180 |
|
Temp PBand |
6.0 |
Temp Int |
30.0 |
Temp Der |
3.0 |
Emissivity |
0.95 |
4181 |
|
Temp PBand |
10.0 |
Temp Int |
30.0 |
Temp Der |
5.0 |
Emissivity |
0.95 |
9142 |
|
Temp PBand |
2.0 |
Temp Int |
30.0 |
Temp Der |
1.0 |
9143 |
|
Temp PBand |
18.0 |
Temp Int |
90.0 |
Temp Der |
30.0 |
9144 |
|
Temp PBand |
18.0 |
Temp Int |
90.0 |
Temp Der |
30.0 |
9170 |
|
Temp PBand |
1.5 |
Temp Int |
20.0 |
Temp Der |
1.5 |
Grad PBand |
5.0 |
Grad Int |
40.0 |
9171 |
|
Temp PBand |
3.0 |
Temp Int |
40.0 |
Temp Der |
1.5 |
Grad PBand |
10.0 |
Grad Int |
80.0 |
9172 |
|
Temp PBand |
12.0 |
Temp Int |
90.0 |
Temp Der |
10.0 |
Grad PBand |
6.0 |
Grad Int |
50.0 |
Grad Ratio |
1.4 |
9173 |
|
Temp PBand |
6.0 |
Temp Int |
100.0 |
Temp Der |
10.0 |
Grad PBand |
10.0 |
Grad Int |
150.0 |
Grad Ratio |
1.3 |
9190A |
|
Temp PBand |
10.0 |
Temp Int |
100.0 |
Temp Der |
8.0 |
Grad PBand |
20.0 |
Grad Int |
150.0 |