Tech Notes
--------------------------------------------------------------------------------------------------------------------------------
Title: 9150 displays Error 7
Created: 6 Apr 2012
Last Revised: 6 Apr 2012
--------------------------------------------------------------------------------------------------------------------------------
Applies To: 9150 Furnaces
Problem Description: Error 7 is displayed on the 9150.
Resolution/Work Around: Error 7 is displayed when the 9150 goes into a hard Cut-Out.
Follow these steps to make sure the cutout is set properly:
- In the secondary functions, under Configuration menu, Operating Parameters menu, make sure the soft cutout, SCO, is set to 1225°C.
- If it is not, set the SCO to 1225.
- If it is, then check the hard cutout.
- In the secondary functions, under Configuration menu, under Calibration menu, CAL, the hard cutout is displayed and should be between 1260°C to 1280°C.
- If it is not, you can adjust the VR1 on the analog board to 1280°C.
- Turn off the power and lay the 9150 on the right side (looking from the front).
- Remove the left side panel. 2 screws on the front, 2 on the back, and 1 on the handle rod.
- Turn on the power and get back to the CAL menu with hard cutout.
- Now, adjust VR1 to 1280°C. Then, put the unit back together.
- After adjusting, cycle the power, the unit should no longer display Error 7.If the use continues to display Error 7, the cutout thermocouple or the analog board will need replaced.