Here are the steps to change the communication interface from the default RS-232 setting to IEEE/GPIB with or without an address change:

Power on the unit.  Connect your IEEE connector to the correct rear port.

1 - Press  the Shift button. The display should now show Shift Enabled.

2 - Press the Display button (4).

3 - Press the Enter button until Remote Interface RS232 is displayed.

4 - Press the down arrow to change to GPIB.

5 - Press the Enter button to go to GPIB Address.

6 - Use the up and down arrows to change the address.

7 - Push Enter to get back to the main display and save changes. 

You should be able to do an instrument scan now and "see" the GPIB instrument in NI MAX.