As a general rule, temperature measurement devices should be regularly calibrated in order to prove their accuracy. This guide to Field Calibration will assist you in the calibration of your device.

Comark has a number of Pocket Digital and Pocket Dial Thermometers that can be calibrated to ensure accurate temperature readings. The goal of the calibration is to verify that the displayed reading you are getting is an accurate one and for such devices a simple crushed ice solution can be used, to calibrate the thermometer to zero degrees Celsius, or 32 degrees Fahrenheit.

Pocket Dial Thermometers

Applicable Products:

T125 | T160A | T160/3 | T200L | T200LC | T220A | T220/3 | T220/38A | T550AK | T550/38A | TC100A | TCF220/3 | TCF220/3K

Patented Nut Method

To re-calibrate to within ± 1 division, immerse the stem at least 3″ into a slush ice solution (60% crushed ice and 40% tap water). Wait at least one minute for the sensor to reach equilibrium. By holding the unique, hand calibrating nut in one hand, the Thermometer can be adjusted until the pointer reads 32°F, as shown below.

Regular Nut Method

To re-calibrate to within ± 1 division, immerse the stem at least 7.5cm / 3″ into a slush ice solution (60% crushed ice and 40% tap water). Wait at least one minute for the sensor to reach equilibrium. By holding the hex nut securely with a wrench or pliers, the Thermometer can be adjusted until the pointer reads the desired temperature. You will need a second Thermometer, either glass or digital, to verify the control temperature.

Pocket Digital Thermometers

Applicable Products:

300| 300B | DT400 | KM14 | KM400AP | PDT300 | PDQ400

Most Comark Digital Thermometers can be Field Calibrated using the simple steps outlined below. We would also encourage you to watch the ‘Slush Ice Thermometer Calibration’ Video if this is your first time re-calibrating your device, or in case you need a quick visual reminder. Click Here to watch.

  • Put some water and crushed ice into a clean mug or water bath (approx., 60% crushed ice to 40% water).
  • Insert the thermometer into the ice solution until the stem is covered to at least halfway along its length.
  • Wait for the reading to settle.
  • Press and hold the D-H button for 8 seconds to begin calibration.
  • “CAL” will be displayed on the LCD screen and calibration at 0°C will be complete.

Note: CAL will be displayed any time the D-H key is pressed for 8 seconds, but the thermometer will not be re-calibrated unless the tip is inserted in ice water

The following Pocket Digital Thermometers all have a built-in calibration function allowing for Field calibration as described above:
 

  • 300 (3058293) – Digital Temperature Probe
  • 300B (3058333) – Digital Temperature Probe with Boot
  • DT400 (3305019) – Pocket Thermometer with Thin Tip
  • KM14 (3851824) – Dishwasher Thermometer
  • KM400AP (4622257) – Waterproof Allergen Thermometer
  • PDT300 (3060281) – Waterproof Pocket Digital Thermometer
  • PDQ400 (3807774) – Waterproof Pocket Digital Thermometer            

Crushed Ice Solution

A simple crushed ice solution can be used, to calibrate your thermometer to zero degrees Celsius, or 32 degrees Fahrenheit. A crushed ice solution typically contains approximately 60% crushed ice to 40% water. We would encourage you to watch the ‘Slush Ice Thermometer Calibration’ video, for additional guidance. Click Here to watch.

First, you will need a suitable container A 250 millilitre, or eight ounce container, would be ideal.

Next you will need to crush enough ice to fill the container. Don’t be tempted to use un-crushed ice, such as standard ice cubes. Once the container is full, pour in cold water until the surface starts to turn from ice white to slightly grey in colour.

The ice on the surface should show just a small amount of water and not an excessive amount, so push down firmly on the ice and pour off any excess. Essentially the ice should still be allowed to float.

Your Ice Point is now ready for use and can be topped up as necessary, depending on how long it’s needed. Simply add more ice but remember to drain any excess water.

To start the calibration check, simply immerse the probe of your thermometer into the solution, so that it is at least 2.5 centimetres, or 1 inch, below the level of the water in the crushed ice solution.

You should not allow the probe tip to rest on the bottom or the sides of the container.

Now you need to wait for the reading on your thermometer, to become stable. We would normally expect a reading to show, that is within 1 degree of zero degrees Celsius, or within 2 degrees of 32 degrees Fahrenheit.