Views:

Download all test results from your Versiv 2 main to your PC (for safety) 
Click HERE for the link to download firmware version 6.14 Build 9 for Versiv 2.
Important Notes:   
Know your PIN code, 6.14 will prompt for PIN when Versiv 2 reboots after the update. 
It is always recommended to plug your charger into the Versiv 2 mainframe before doing a software update. Loss of power due to a discharged battery during a software update can put your Versiv 2 unit into an unusable state.  
Even though you are performing a field downgrade on the firmware version, all actions related to this on the Versiv 2 mainframe and in Linkware PC refer to it as “update.” 
Versiv 2 Downgrade Procedure
Put the Versiv 2 in rescue mode
For Versiv 2 Mainframe:

  • Power off the unit. 
  • Hold down the HOME and TEST keys. 
  • Keeping them held down, turn on the tester, then release the Home and Test keys. 
  • You will see a "RESCUE MODE" screen appear. 

Once in rescue mode, run the latest version of LinkWare PC, connect to Versiv 2 over USB, and select the function: UTILITIES → DSX CABLEANALYZER → SOFTWARE UPDATE

Linkware PC will start the process of downgrading. Do not touch the Versiv 2 main unit till it finishes this process. It will restart automatically in normal mode after the downgrade is finished. 
After the update/reboot of Versiv 2, check with Linkware PC the new version of firmware in the main unit by selecting: UTILITIES → DSX CABLEANALYZER → SOFTWARE VERSION. 
You should get a window confirming that your main instrument has firmware version 6.14 Build 9. 
Final Steps
Dismiss any dialog boxes to update the Versiv remote, set a reference between Versiv 2 main and Versiv remote, and continue on with your testing. 
If the above procedure was unsuccessful, the next step would require you to contact your local service center for repair.

.