Applies To: All large refrigerated baths
Problem Description:
In an effort to make Fluke products RoHS compliant we have changed from the old manually adjusted version of this line monitor, that was not RoHS compliant, to a new digitally adjusted version, that is RoHS compliant.
Resolution/Work Around:
The settings should be as follows:
Setup
Voltage: 115/230 (depending on model)
Time Delay: 120 seconds
Over/under voltage: 12%
Control mode: off
Response time: 5 seconds
Fault 1-5: [ignore these, they can be anything, it’s just a record of previous cutouts]