Title: Factory PID settings for my Metrology Well                                           

Created:  9 Jan 2013

Last Revised:

--------------------------------------------------------------------------------------------------------------------------------

 

 

Applies To:  Various Metrology Wells

 

Problem Description:  Unit will not maintain or reach temperature. PID is wrong.

 

Resolution/Work Around:   Set the PID back to factory default.

 

4180

Temp PBand

6.0

Temp Int

30.0

Temp Der

3.0

Emissivity

0.95

 

4181

Temp PBand

10.0

Temp Int

30.0

Temp Der

5.0

Emissivity

0.95

 

9142

Temp PBand

2.0

Temp Int

30.0

Temp Der

1.0

 

9143

Temp PBand

18.0

Temp Int

90.0

Temp Der

30.0

 

9144

Temp PBand

18.0

Temp Int

90.0

Temp Der

30.0

 

9170

Temp PBand

1.5

Temp Int

20.0

Temp Der

1.5

Grad PBand

5.0

Grad Int

40.0

 

9171

Temp PBand

3.0

Temp Int

40.0

Temp Der

1.5

Grad PBand

10.0

Grad Int

80.0

 

9172

Temp PBand

12.0

Temp Int

90.0

Temp Der

10.0

Grad PBand

6.0

Grad Int

50.0

Grad Ratio

1.4

 

9173

Temp PBand

6.0

Temp Int

100.0

Temp Der

10.0

Grad PBand

10.0

Grad Int

150.0

Grad Ratio

1.3

 

9190A

Temp PBand

10.0

Temp Int

100.0

Temp Der

8.0

Grad PBand

20.0

Grad Int

150.0