Views:

The CPC600 modules can be calibrated and adjusted by COMPASS for Pressure when using the Cal Sled accessory.  Both the Primary and Secondary module ranges are supported.  The process involves the following steps:

  1. collect As Found data
  2. evaluate if the module is in tolerance, marginally in tolerance, or out of tolerance
  3. determine and apply a straight line best fit to correct the module
  4. collect As Left verification data

The attached tutorial provides guidance on the key configuration settings for the DUT Definition and the Test Definition, and it includes the VB Script code for the Test Macros.

 

Related Attachments (1)